Abstract

Dear Editor:
Five commercially available devices were examined: Paradigm® Veo™ (Medtronic Minimed, Minneapolis, MN), D-Tron, Spirit Combo, Dana (Sooil, Seoul, Korea), and OmniPod® (Insulet Corp., Bedford, MA). The catheter was occluded completely applying a clip at the tip of the catheter at t=0. Two commonly used CSII delivery rates were chosen: 1.0 IU/h and 0.5 IU/h. Insulin aspart was used in all pumps. The fluid pressure in the catheter depends on its compliance and thereby on its length, and therefore two different lengths of the catheter were tested: 60 cm and 2.5 cm. The OmniPod was only tested with a length of 2.5 cm. All tests were performed in duplicate except for the OmniPod, for which testing was done three times at a basal rate of 1.0 IU/h and three times at 0.5 IU/h. It was assured that no insulin came out of the catheter during the observation time.
The results are given in Table 1. Time to occlusion alert varied from 1 h 52 min to 4 h 32 min at a basal rate of 1.0 IU/h with infusion catheters of 60 cm. At a basal rate of 0.5 IU/h the occlusion alert time varied from 1 h 34 min to 9 h 25 min. No large differences were seen between basal rates in the Paradigm Veo and Spirit Combo. If the length of the infusion set was 60 cm with a basal rate of 0.5 IU/h, the time to occlusion alert doubled in Dana and almost tripled in D-Tron compared with a basal rate of 1.0 IU/h. The alert in the Dana pump has three sensitivity levels: low, medium, and high. Increasing the sensitivity from medium to high decreased the time until an alert was given. If the infusion set was occluded at 2.5 cm with a basal rate of 1.0 IU/h, the occlusion alert occurred faster in three of four insulin pumps compared with 60 cm. If the insulin rate was 0.5 IU/h, two devices had an earlier alert with an occlusion of 2.5 cm compared with the infusion set of 60 cm. Overall, the OmniPod seemed to need more time to activate the occlusion alert than the other pumps.
In conclusion, the time until occlusion alert given in the manufacturers' manuals was somewhat longer than we found in our study. In three of the four devices the time until occlusion alert at a catheter length of 2.5 cm was shorter than with a catheter length of 60 cm, if the basal rate was 1.0 IU/h. This is in accordance with the fact that the distention capacity of the 2.5-cm catheter is less than that of the 60-cm catheter. However, it remains unclear why this was not seen in the fourth pump. At a basal rate of 0.5 IU/h, the time until occlusion alert at a catheter length of 2.5 cm was shorter in two of the four devices compared with 60 cm. An explanation for our finding that the alert did not occur faster in the other two devices despite the shorter catheter length could be that the displacements of the lead screw are too small to build a threshold pressure to trigger the alert at this low basal rate. The dislocations of the screw lead are equivalent to the step sizes of insulin infusion. In the two devices with an earlier alert, the step size is bigger than in the other two devices: 0.1 compared with 0.05 IU of insulin. Therefore, in these devices, the threshold pressure is built up earlier and the occlusion is detected faster.
Thus the time until an occlusion alert differed per basal rate, length of the infusion catheter, and the commercial device, although the pattern is not always completely predictable or consistent. Efforts from manufacturers to decrease the time between occlusion and alarm would be welcome. For the time being patients and caregivers should realize that occlusion alerts are not very sensitive.
