Abstract

Introduction
S
The themes of the literature for SMBG have not changed. Accuracy remains a major concern, but the traditional name brand meters seem to do quite well. Alternate site testing, use of SMBG for telemedicine, inpatient SMBG, and the use of SMBG in the pediatric age group continues to be addressed, although not in the volume we saw 10 years ago. We hope you find this review of the past year in SMBG helpful.
Key Articles Reviewed for this Article
Perioperative glucose monitoring and treatment of patients undergoing vascular surgery in a community hospital setting
Morrison S1, O'Donnell J2, Ren D3, Henker R4
Can gingival crevicular blood be relied upon for assessment of blood glucose level?
Dwivedi S1, Verma SJ2, Shah M3, Jain K1
Internet blood glucose monitoring systems provide lasting glycemic benefit in type 1 and 2 diabetes: a systematic review
Tildesley HD1, Po MD2, Ross SA3
Perceptions and experiences of using automated bolus advisors amongst people with type 1 diabetes: a longitudinal qualitative investigation
Lawton J1, Kirkham J1, Rankin D1, Barnard K2, Cooper CL3, Taylor C4, Heller S5, Elliott J5, the REPOSE Group
Management of endocrine disease. Effects of telecare intervention on glycemic control in type 2 diabetes: a systematic review and meta-analysis of randomized controlled trials
Huang Z, Tao H, Meng Q, Jing L
Testing versus guessing blood glucose values: impact on self-care behaviors in type 2 diabetes
Pettus J1, Stenger P2, Schachner HC3, Dunne N3, Parkes JL3, Pardo S3, Edelman SV1,4
Accuracy evaluation of blood glucose monitoring systems in children on overnight closed-loop control
DeSalvo DJ, Shanmugham S, Ly TT, Wilson DM, Buckingham BA
Frequency of mealtime insulin bolus predicts glycated hemoglobin in youths with type 1 diabetes
Patton SR1, DeLurgio SA2, Fridlington A3, Cohoon C3, Turpin AL3, Clements MA3
Deciding between using the first or second drop of blood for the self-monitoring of blood glucose
Li M, Wang X, Shan Z
System accuracy evaluation of systems for point-of-care testing of blood glucose: a comparison of a patient-use system with six professional-use systems
Freckmann G1, Schmid C1, Pleus S1, Baumstark A1, Link M1, Stolberg E1, Haug C1, Sieber J2
Perioperative glucose monitoring and treatment of patients undergoing vascular surgery in a community hospital setting
Studies in the past have shown that the glucose control of patients that underwent cardiovascular surgery directly affected the outcome of the surgery: they stayed longer at the hospital, with incidence of infection, stroke, renal failure, and myocardial infarction. Some were readmitted to the hospital within 30 days of their discharge. This article discusses the glucose monitoring and treatment practices of a small community hospital, when its goal is to determine how these practices relate to postoperative complications in patients undergoing vascular surgery. This was a retrospective analysis of 101 patients who underwent vascular surgery and whose charts were reviewed. Details of perioperative blood glucose levels, frequency of monitoring, treatment practices, and postoperative complications were collected from their electronic and paper medical records. Twenty-seven had a blood glucose level greater than 140 mg/dL preoperatively. Intraoperative blood glucose levels were monitored for eight patients. Nine patients had blood glucose levels treated during the perioperative period. The threshold for treatment of preoperative hyperglycemia was a blood glucose level of 236 mg/dL. Increased attention of perioperative blood glucose levels is needed so that appropriate interventions can be established for the improvement of outcomes. Glucometers must be available to anesthesia providers in order to perform intraoperative monitoring of blood glucose levels.
The debate on intensive versus less intensive diabetes management during the perioperative period continues. The investigators treated preoperative hyperglycemia at a level of 236 mg/dL. There is no justification for using such a high number for initiating treatment for hyperglycemia. However, the authors emphasize an important point that glucose meters must be readily available to anesthesiologists so that appropriate peri- and intra-operative monitoring of blood glucose can be performed. The major reason for ongoing controversy revolves around the risk of hypoglycemia in this high-risk population in the hospital setting. In addition, a majority of the hospitals still do not have standardized protocols to manage hyperglycemia in the perioperative setting.
Can gingival crevicular blood be relied upon for assessment of blood glucose level?
Approximately half of the patients suffering from diabetes mellitus (DM) go undiagnosed. In addition, the prevalence of DM is more than twice as high as in patients with periodontitis when compared to periodontally healthy subjects. Therefore, a high percentage of patients with periodontitis may have undiagnosed DM. The aim of the present study was to determine whether the blood from a gingival crevice during a routine periodontal examination can indicate glucose levels.
Material and Methods
In total, 75 patients (43 males and 32 females) with chronic periodontitis were divided into two groups—Group I and Group II, respectively—for observational cross-sectional studies. Blood oozing from the gingival crevices of anterior teeth following periodontal probing was collected with the stick of a glucose self-monitoring device, and the blood glucose levels were measured. At the same time, finger-prick blood was taken for glucometric analysis and subsequent readings were recorded.
Results
The patient's blood glucose values ranged from 74 to 256 mg/dL. The comparison between gingival crevicular blood and finger-prick blood showed a very strong correlation, with a t-value of 3.97 (at P-value=0.001).
Conclusions
The data from this study have shown that gingival crevicular blood (GCB) collected during diagnostic periodontal examination can be an excellent source of blood for glucometric analysis.
Alternate sources of blood continue to be of interest for glucose monitoring in high-risk patients. This study performed in patients with periodontitis found a good correlation of GCB with finger-prick blood glucose values. More importantly, this study highlighted undiagnosed diabetes in patients with periodontitis. The blood glucose measurements correlated well across different glucose ranges from 74 to 256 mg/dL. It is unknown if the rate of change of blood glucose values will be affected at the lower and the higher glucose level since those levels are not available in this study.
Internet blood glucose monitoring systems provide lasting glycemic benefit in type 1 and 2 diabetes: a systematic review
Internet blood glucose monitoring systems (IBGMS) have been shown to improve glycemic control in patients with type 2 diabetes (T2D) who are pharmacologically managed, using oral agents or insulin. IBGMS improves glycemic levels in patients with type 1 diabetes (T1D) and does not lead to increased hypoglycemia. Mechanisms underlying IBGMS-associated glycemic improvement go beyond optimizing insulin dose titration. The most relevant effects have to do with increased self-motivation of patients and improved communication between patients and physicians. IBGMS have been recommended in clinical practice guidelines, and their effectiveness and safety in trials suggest that this approach is appropriate for patients with T1D or T2D.
This study highlights the role of IBGMS. The study documented improved glucose control in patients with type 2 diabetes managed either with oral drugs or insulin treatment. The most important lesson from the study is improved patients' self-motivation and better patient/physician communication. It has been highlighted in several studies that patient engagement and access to the provider results in better outcomes. Electronic communication makes such interactions easier for patients and providers. Longer studies for patients with type 1 and type 2 diabetes will need to be performed before concluding effectiveness of IBGMS.
Perceptions and experiences of using automated bolus advisors amongst people with type 1 diabetes: a longitudinal qualitative investigation
Aims
To explore people's reasons for using bolus advisors to determine insulin doses, as well as their experiences using bolus advisors and their likes/dislikes of this technology.
Subjects and Methods
A total of 42 people with type 1 diabetes who had received instructions on using bolus advisors during a structured education course were interviewed post-course and 6 months later. Data were analyzed thematically.
Results
Participants who considered themselves to have poor mathematical skills showed a gratitude for, and heavy reliance on, advisors. Others liked and chose to use advisors because they saved time and effort calculating doses and/or had a data storage facility. Follow-up interviews reflected that, by virtue of no longer calculating their doses, participants could become deskilled and increasingly dependent on advisors. Some forgot what their mealtime ratios were; others reported a misperception that, because they were preprogrammed during courses, the parameters never needed changing. Effective review of blood glucose data could be hindered by the use of data storage facilities, and some participants reported an adverse impact on glycemic control.
Discussion
While participants liked advisors and perceived benefits to using them, there may be unintended consequences to giving people access to this technology. To promote effective use, ongoing input and education from trained health professionals might be warranted.
Patients using MDI or insulin pump increasingly use bolus advisors. Many times bolus advisors take into account insulin on board and calculate the appropriate bolus insulin dose for the meal based on carbohydrate intake. The main point highlighted in this study is patient education by the health provider team to avoid unintended consequences. In addition, it's important that patients do not expect more than what the advisor can provide, as some patients reported adverse effect on their glucose control.
Management of endocrine disease. Effects of telecare intervention on glycemic control in type 2 diabetes: a systematic review and meta-analysis of randomized controlled trials
Objective
To review the published literature on the effects of telecare intervention in patients with type 2 diabetes as well as inadequate glycemic control.
Design and Methods
From December 8–16, 2013, we conducted a review of randomized controlled trials on telecare intervention in patients with type 2 diabetes, along with a search of electronic databases such as The Cochrane Library, PubMed, EBSCO, CINAHL, Science Direct, Journal of Telemedicine and Telecare, and China National Knowledge Infrastructure (CNKI). The eligible studies were independently selected and reviewed by two evaluators. The following elements were analyzed: changes in HbA1c, fasting plasma glucose (FPG), post-prandial plasma glucose (PPG), body mass index (BMI), and body weight.
Results
The analysis included 18 studies and 3,798 subjects. It showed that telecare greatly improved the management of diabetes. Mean HbA1c values were reduced by −0.54 (95% CI, −0.75 to −0.34; P<0.05), mean FPG levels by −9.00 mg/dL (95% CI, −17.36 to −0.64; P=0.03), and mean PPG levels reduced by −52.86 mg/dL (95% CI, −77.13 to −28.58; P<0.05) when compared with the group receiving standard care. Meta-regression and subgroup analyses indicated that study location, sample size, and treatment-monitoring techniques were the sources of heterogeneity.
Conclusions
Patients monitored by telecare showed significant improvement in glycemic control in type 2 diabetes when compared with those monitored by routine follow-up. Significant reduction in HbA1c levels was associated with Asian populations, small sample size, and telecare, and with those patients with baseline HbA1c greater than 8.0%.
The global increase in the prevalence of diabetes is likely to affect more than half a billion people over the next decade, especially in countries with limited resources. In addition, lack of providers in the field of diabetes mandates development of telecommunication intervention for management of diabetes. This is a meta-analysis of 18 different studies (Asian population) that concluded that significant improvement in A1c (−0.54%) with significant improvement in fasting and postprandial blood glucose levels when compared with the standard care. Telecare also allows reach out to remote populations where limited care is available. It is important that payers should recognize telecare as an alternate way to deliver a similar level of diabetes care as seen in routine clinical practice. Many parts of the world, Telecare is not reimbursed at the same level or at all.
Testing versus guessing blood glucose values: impact on self-care behaviors in type 2 diabetes
Objective
To assess differences between estimated blood glucose values and those measured on a blood glucose meter and the impact on self-care behavior in type 2 diabetes.
Design and Methods
Subjects ≥18 years with type 2 diabetes (N=297) attending a Taking Control of Your Diabetes conference were asked questions about diabetes management and to estimate their current blood glucose. Study staff tested subjects' blood glucose on a meter. After seeing the result, subjects were again asked questions on diabetes management.
Main outcome measure
The percentage of subject blood glucose estimations that were outside ISO 15197:2003 accuracy criteria (>±15 mg/dL or >±20% of meter glucose values).
Results
Nearly half (46%) of subjects estimated blood glucose values outside ISO 15197:2003 accuracy criteria. Time since last blood glucose test, time since last meal, testing frequency, and A1C did not have an effect on differences between estimated blood glucose values and meter results. In the questionnaire before blood glucose testing, most subjects strongly agreed, agreed, or neither agreed nor disagreed that “I make decisions about my diabetes, such as my food intake or my insulin dose even when I do not test my blood sugar” (71%) and “My body tells me without testing if my blood sugar is low or high” (77%). After blood glucose testing, 99% of subjects strongly agreed, agreed, or neither agreed nor disagreed that “Knowing my blood sugar by checking could help me make different diabetes decisions.”
Conclusions
Self-monitoring of blood glucose is an important component of diabetes self-management. Testing rather than guessing blood glucose values is important to obtain accurate results and inform people with type 2 diabetes to make effective, appropriate diabetes management decisions. A potential limitation of this study is that the subject population may not be representative of the general population of people with diabetes; however, the conference setting may attract a more motivated population, which could underestimate the magnitude of the results.
A common defense for many patients why they don't measure blood glucose is they think they can “feel” their glucose levels without testing. This is clearly not the case. While the authors acknowledge the limitations of the study population, my experience is these findings would be true for most patients.
Accuracy evaluation of blood glucose monitoring systems in children on overnight closed-loop control
In this pilot study, children with type 1 diabetes participating in overnight closed-loop studies were evaluated to determine the difference in accuracy between the Bayer Contour® Next (CN) and HemoCue® (HC) glucose monitoring systems. A total of 28 subjects aged 10–18 years old were admitted to a clinical research center and glucose values were obtained every 30 minutes overnight. Using whole blood samples for CN and HC readings, glucose values were measured and results were compared to Yellow Springs Instrument (YSI) reference values obtained with plasma from the same sample. System accuracy was compared using mean absolute relative difference (MARD) and International Organization for Standardization (ISO) accuracy standards. Glucose measurements were evaluated at 457 time points. CN performed better than HC with an average MARD of 3.13% compared to 10.73% for HC (P<0.001). With a limited sample size, CN met ISO criteria (2003 and 2013) at all glucose ranges while HC did not. CN performed very well, and would make an excellent meter for future closed-loop studies outside of a research center.
One unexpected benefit from the closed-loop studies has been the ability to analyze accuracy of both continuous glucose meters and home blood glucose monitors. This study is surprising, despite the relatively few patients, in that the home meter has a lower MARD than the HemoCue®, which has traditionally been a gold standard for home studies. This further supports the improved technology for today's home glucose meters, and makes us wonder why we need to continue to use the more inaccurate meters in the United States for Medicare and indigent patients.
Frequency of mealtime insulin bolus predicts glycated hemoglobin in youths with type 1 diabetes
Background
Two electronic measures of adherence exist within pediatric diabetes management: frequency of daily blood glucose monitoring (BGM) and the BOLUS score, a measure of frequency of mealtime insulin bolusing. Previous research has shown the BOLUS score to be superior to daily BGM in predicting youths' glycated hemoglobin (HbA1c) in a cross-sectional study. Using a prospective, longitudinal design, we present data comparing the two adherence measures in predicting HbA1c.
Subjects and Methods
Blood glucose meter data and insulin pump records were collected from a clinical database of 175 youths with type 1 diabetes (mean age, 11.7±3.6 years at baseline), as were the youths' HbA1c levels occurring at the download time and at 3, 6, 9, and 12 months post-downloads. Using a standardized protocol, the youths' mean BGM and BOLUS score were calculated.
Results
Intraclass correlations (ICCs) revealed significant absolute equivalence among youths' predicted HbA1c values using BOLUS and BGM scores and future actual HbA1c values up to 12 months post-download. However, the ICCs of BOLUS scores with future HbA1c values were consistently higher than those of the BGM scores. Also, the predictions of the BOLUS scores were significantly more accurate (P≤0.002) than those of the BGM scores based on the root-mean-squared error of predictions.
Conclusions
In a prospective, longitudinal design, youths' BOLUS scores were superior to youths' daily BGM in predicting future values of HbA1c. Calculating a BOLUS score versus BGM can help researchers and clinicians better predict the youths' HbA1c.
It is well-appreciated that many pediatric patients miss their boluses for both meals and snacks, and this has been a major challenge for modern-day insulin regimens in this age group. It is not surprising that quantitatively a “BOLUS SCORE” is more important to the HbA1c level than that frequency of blood glucose testing. One has to wonder if this is even more of an issue for multiple-daily injection patients where prandial insulin is not as convenient.
Deciding between using the first or second drop of blood for the self-monitoring of blood glucose
Aims
To evaluate whether the first or the second drop of blood is more effective for the self-monitoring of blood glucose (SMBG).
Methods
Hospitalized patients carried out SMBG using their first and second drop of blood; meanwhile, venous blood glucose was measured. The differences in blood glucose measurements were then compared in groups with varying regions of blood glucose levels. There were 802 groups of blood glucose in 526 patients.
Results
No significant difference was found in the blood glucose levels of the first and second drop of blood and venous blood. However, after combining then dividing measurements into six groups according to blood glucose concentration, we found statistically significant differences in the groups containing blood glucose values <9.9 or 20–30 mmol/L among the blood glucose levels obtained from the first drop, second drop, and venous blood. Yet, there were no significant differences in the 10–14.9 or 15–19.9 mmol/L groups.
Conclusions
Both the first or second drop of blood can be used for performing SMBG to assess real-time venous glucose in the clinical setting. By categorizing blood glucose into various levels more accurately, we observed that there was no significant difference between the first or second drop of blood and the venous blood glucose value when blood glucose levels were maintained between 10 and 20 mmol/L. In addition, when blood glucose levels were below 10 mmol/L, the value obtained from first drop of blood was close to that from venous blood, whereas when the blood glucose level is >20 mmol/L, the blood glucose value from the second drop of blood was more accurate.
This study brings back memories of the “first-versus-second voided urines” from the urine glucose testing days. This study needs to be repeated with different blood glucose meters to confirm the results. From a practical point of view, both at home and in the hospital it seems most use the first drop of blood, but even this isn't well documented. Assuming the first drop is indeed used most of the time, the greatest impact of this would be for hyperglycemia above 20 mmol/L, for which clinically one could argue accuracy is not as critical with insulin dosing as with hypoglycemia.
System accuracy evaluation of systems for point-of-care testing of blood glucose: a comparison of a patient-use system with six professional-use systems
Background
Point-of-care testing (POCT) of blood glucose (BG) is performed by medical personnel in clinical settings as well as by patients at home employing SMBG. Our research aimed to determine if a system mainly intended for SMBG by people with diabetes, but also suitable for BG measurements by medical personnel, can achieve measurement accuracy on capillary blood samples that are comparable to professional-use POCT systems.
Methods
System accuracy was evaluated under standardized conditions, following the ISO standard 15197:2003. For each system (which included one SMBG system with three test strip lots and six professional-use systems), measurement results from capillary blood samples of 100 subjects were compared with a standardized laboratory glucose oxidase method (YSI 2300 glucose analyzer).
Results
The seven evaluated systems showed 99.5% or 100% of the measurement results within the required system accuracy limits of ISO 15197:2003 (±0.83 mmol/L at BG concentrations <4.2 mmol/L and ±20% at BG concentrations ≥4.2 mmol/L). Using the more stringent requirements of the revision ISO 15197:2013, the systems showed between 99% and 100% of the measurement results within the accuracy limits (±0.83 mmol/L at BG concentrations <5.55 mmol/L and ±15% at BG concentrations ≥5.55 mmol/L). In addition, the systems showed between 82% and 98% when even more restrictive limits were used (±0.56 mmol/L and ±10%, respectively).
Conclusions
Data from this study, which aimed to determine system accuracy, suggest that SMBG systems can achieve system accuracy that is comparable with professional-use systems when measurements are carried out on capillary blood samples by trained personnel in a standardized and controlled setting.
While this is one more piece of evidence about the improved accuracy of modern-day blood glucose meters, these measurements were performed “by clinical personnel well-trained to the limitations of the test systems, the manufacturer's labelling, the safety practices, and the study protocol.” A better protocol would have been to have the identical study performed by patients of all ages and hospital nurses involved with both subcutaneous and intravenous insulin dosing.
Author Disclosure Statement
S.G. has no competing financial interests. I. B. H. is a consultant for Roche and Abbott and conducts research for NOVO.