In late 2017, the US Food and Drug Administration (FDA) and Centers for Medicare and Medicaid Services (CMS) jointly approved Foundation Medicine's FoundationOne CDx, a comprehensive diagnostic and genomic profiling assay for solid tumors that is intended to inform the use of targeted oncology therapies. The decision was only the second affirmative determination to be reached via the Parallel Review Pathway, which was launched in 2011 as a means for manufacturers of high-risk medical devices to meet the review requirements of FDA and CMS simultaneously. Under this review process, CMS assesses its national coverage determination concurrently with review of safety and effectiveness data by the FDA. This pathway was designed to build on the agencies' overlapping mandates and “common interest in improving the health of patients through the availability of safe, effective, and affordable medical products,” 1 while enhancing the speed and transparency of decision-making that leads to the sale of new medical devices. But why, if some policymakers and the medical device industry suggest that the US market needs more regulatory efficiency, has parallel review been used so infrequently? We sought to review the initial promise of the parallel review pathway and consider explanations for its unmet expectations.
Approval and Coverage of Medical Devices in the US
In the US, new high-risk medical devices must undergo formal FDA review before introduction to the market. 2 Manufacturers of devices such as coronary stents and implantable cardioverter-defibrillators, intended for life-sustaining or life-supporting purposes, most often submit applications via the Premarket Approval (PMA) process. While specific requirements are tailored to device type, the PMA pathway involves developing preclinical and clinical data that demonstrate “reasonable assurance of safety and effectiveness.” 3 This standard involves sufficient valid scientific evidence to demonstrate that the probable health benefits attributable to the intended use of the device outweigh any probable risks, and that such benefits result in clinically meaningful results. 4 The FDA has historically considered “valid scientific evidence” for the determination of effectiveness to consist of well-controlled clinical investigations, although many high-risk devices are approved based on single-arm trials without active comparators or other less rigorous data sources, with evidentiary standards tailored to the device and clinical context. 5
Once FDA approval is granted, a device sponsor must submit an application to CMS to determine whether the product is “reasonable and necessary,” which are Medicare's statutory coverage criteria. Necessity has been broadly interpreted to reflect the strength and rigor of the clinical evidence supporting a device, taking into account factors including study design and sample size. 6 By contrast, the meaning of “reasonable” has not been clearly defined with regard to medical devices. Some have postulated that this refers to devices that have sufficient evidence to support clinically-relevant improved health outcomes, 7 achieved through assessment of clinical data for study quality, relevance to Medicare beneficiaries, and magnitude of benefits and risks ( Table 1 ). 8
Evidence Reviewed to Determine a Medical Device “Reasonable and Necessary” for National Coverage Determination as Delineated in CMS Decision Memos
FDG-PET: fluorodeoxyglucose positron emission tomography; FDA: Food and Drug Administration; NICE: National Institute for Health and Care Excellence; PARTNER: Placement of Aortic Transcatheter Valves trial; ACCF: American College of Cardiology Foundation; AATS: American Association for Thoracic Surgery; SCAI: The Society for Cardiovascular Angiography and Interventions; STS: Society of Thoracic Surgeons.
Responsibility for making the reasonable and necessary determination for the vast majority of devices falls to fiscal intermediaries that serve as representatives of different Medicare districts within the US. These contractors assess whether a device meets the reasonable and necessary criteria, resulting in a local coverage determination. Local coverage decisions may limit the coverage of items or services to specific diagnoses, or may preclude coverage entirely, with each decision applicable within the contractor's local jurisdiction. 9
Alternatively, manufacturers, clinicians, or CMS itself may request a national coverage determination (NCD), which grants, limits, or excludes Medicare coverage nationwide, particularly when significant uncertainty regarding the risks and benefits of a new device exists or when non-uniform local coverage determinations exist. NCDs are issued on approximately 5-15 devices or services annually, and are targeted at those posing special concerns related to cost, safety, or effectiveness among Medicare beneficiaries. Until 2017, NCDs were published prior to implementation and local coverage decisions were not publicly available. However, under the 21st Century Cures Act, fiscal intermediaries were also required to publish the full local coverage decision, public comments and responses, a summary of the evidence considered, and the rationale for the determination, thereby clarifying the data upon which these determinations are decided. 10
In practice, most applications for NCDs are approved. Among NCDs made by CMS from 1999-2007, a favorable decision was reached in more than 60% of the cases. 11 Another potential outcome could be issuance by CMS of “coverage with evidence development,” in which reimbursement for the device is provided only in the context of the collection of additional clinical data, such as through registries or clinical investigations. 12 Approximately two dozen devices or services are currently subject to coverage with evidence development decisions stretching back to 2006. 13
The existence of separate formal review processes for high-risk medical devices at FDA and CMS has contributed to worries about the dissemination of new products and the possibility of delays in timely delivery of devices to market. 14 The independent CMS process may introduce delays between FDA approval and CMS reimbursement for the medical device's utilization. 15 An analysis of the six NCDs issued by CMS in 2013 found that the average time from initiation of analysis to release of the proposed decision was approximately 6 months, with an additional 3 months from proposed to final decisions, and another 4 from posting of the final decision to publishing of implementation instructions. 16 For example, the WATCHMAN device for left atrial appendage exclusion was FDA-approved in March of 2015, but only received an NCD from CMS in April of 2016. Though comparable data are not available regarding implementation of local coverage determinations, their timeline is generally thought to be more expeditious than that of national coverage. 17
The Parallel Review Process and its Underuse
In June of 2010, the FDA and CMS signed a memorandum of understanding to work together “to promote initiatives related to the review and use of FDA-regulated drugs, biologics, medical devices, and foods” with the express purpose to “promote collaboration and enhance knowledge and efficiency by providing for the sharing of information and expertise between the Federal partners.” 18 By 2011, the parallel review pilot program was launched between the FDA's Center for Devices and Radiological Health (CDRH) and CMS. The program was designed not to alter the review criteria for either agency, but to shorten the review process and promote transparency and efficiency. In particular, earlier CMS involvement was thought to encourage clinical trial design that would promote data collection to serve the evidentiary requirements of the FDA and CMS, with a particular focus on therapeutics likely to impact Medicare beneficiaries.
To meet appropriateness criteria for consideration of participation in the parallel review pathway, the device manufacturer must (1) agree to meet jointly with FDA and CMS prior to initiating clinical investigation to incorporate feedback from both parties regarding trial design and analysis, (2) document that the device requires an application for premarket approval (or a petition for de novo review, which classifies a novel device that would otherwise be high-risk to a lower risk tier), and (3) demonstrate that the device is not statutorily excluded from Medicare's hospital and general medical benefit categories. 19
By 2016, the parallel review process received 60 inquiries and 26 applications, with the program purporting to possess the resources to undertake review of 3-5 devices per year. However, since the program's announcement in 2011, only four manufacturers have publicly announced their participation in the dual-review process, with two affirmative, one negative, and one pending determination ( Table 2 ). Reasons why not all 26 devices have been accepted by FDA and CMS for parallel review include existence of a prior NCD or ineligibility for an NCD determination. However, the details of applications and reasons for denial are not publicly available.
Medical Devices Evaluated via the Parallel Review Pathway
S-ICD: Subcutaneous Implantable Cardioverter Defibrillator; MADIT S-ICD: Multicenter Automatic Defibrillator Implantation Trial with Subcutaneous Implantable Cardioverter Defibrillator
Several factors may explain why parallel review has received limited interest. First, the lack of utilization of the parallel review process likely reflects manufacturers' unwillingness to undertake the risk associated with seeking an NCD. 20 In practice, most FDA-approved devices are covered through local coverage decisions. Thus, the benefits of expedited progression through FDA and CMS review may not be worth the risks of a negative determination that precludes or markedly limits reimbursement nationally for Medicare patients. 21 Similarly, increased scrutiny from CMS may lead to additional post-market requirements (such as a mandatory registry), which may not be mandated by local administrative contractors, adding another layer of potential financial risk for sponsors. That said, it is unlikely that the fear of a restrictive NCD accounts entirely for low participation in the parallel pathway, as most devices that do go through the NCD process end up favorably reviewed.
Second, manufacturers often do not design device trials that demonstrate effectiveness in populations specifically relevant to CMS, which will necessarily include older patients who are more likely to have multiple comorbid conditions. 22 For example, a study of 82 studies published in 2015 supporting the approval of 42 high risk medical devices revealed that only 9% of trials analyzed effectiveness by age and that results by age were reported on only 7% of device labels. 23 From a clinical and policy perspective, increasing testing of new products in such patients is an important public health goal because many of these patients will end up receiving approved high-risk devices. 24 Yet evaluating safety and effectiveness endpoints in this population may complicate outcomes ascertainment and increase safety concerns about devices since such patients are likely to be at greater risk for adverse events. 25
Despite the strong rationale for parallel review, this pathway has not been used widely. Strengthening the incentives for sponsors to pursue parallel review may require new policy approaches that either increase the eligible device submissions, or shift the burden of proof regarding clinical effectiveness in Medicare beneficiaries more explicitly towards manufacturers.
From a device manufacturer's standpoint, it is more advantageous — and less expensive — to demonstrate clinical improvements in a healthier and more homogenous younger population, and if necessary argue afterward for extrapolation to Medicare beneficiaries. For example, Medicare currently covers cardiac resynchronization therapy for patients with systolic heart failure and clinical characteristics aligned with pivotal clinical trials, despite its own technology assessment acknowledging the lack of trials focused on Medicare-eligible patients and older (>75) patients in particular. 26 But this permissive approach may not serve either patients' or CMS' interests well, as it is not at all clear that cardiac resynchronization therapy (particularly when paired with defibrillator back-up) actually improves outcomes meaningfully for older patients. 27 If CMS, working with manufacturers early in the investigative process, had made it clear that high-quality evidence specific to older patients would be necessary for coverage of cardiac resynchronization therapy, clinical trials might have been redesigned accordingly. Ideally, parallel review would encourage exactly these study refinements rather than a calculated assumption that a new device or labeling will not be met with sufficient skepticism to merit a restrictive NCD.
Future Directions
Despite the strong rationale for parallel review, this pathway has not been used widely. Strengthening the incentives for sponsors to pursue parallel review may require new policy approaches that either increase the eligible device submissions, or shift the burden of proof regarding clinical effectiveness in Medicare beneficiaries more explicitly towards manufacturers.
Increasing eligibility for the program may be simpler. While initially conceived to focus on high-risk devices subject to PMA review, parallel review could be expanded to include selected medium-risk devices. Most medium-risk devices are evaluated through the 510k pathway, through which devices are evaluated based on whether they introduce new safety and effectiveness concerns compared to a predicate device already on the market to which they are compared. 28 Notably, the 510k pathway has been criticized for inadequately evaluating new devices and has even been recommended for replacement by a National Academy of Medicine panel, in part because most 510k applications do not include clinical data. 29 At the same time, discordance between FDA and CMS decisions may be more common for 510k-cleared devices, 30 suggesting an opportunity for sponsors to incorporate joint feedback from FDA and CMS in application design. Pairing requirements to include more routine collection of clinical data prior to marketing select medium-risk devices with making these submissions eligible for parallel review might partially address criticism of the 510k pathway while also increasing program utilization. For example, it has been suggested that medium-risk device types such as molecular diagnostics may be particularly well-suited to parallel review because of comparatively slow clearance. 31 Newer technology subject to 510k clearance, such as selected mobile applications, might also be targeted for inclusion in parallel review as a means towards encouraging CMS to provide payment for services — such as review of arrhythmia recordings sent directly from patients to their providers — that currently are not eligible for reimbursement.
Alternatively, CMS might motivate program participation by adopting a more aggressive stance regarding requirements for data specific to Medicare beneficiaries. These patients have been shown to differ substantially in terms of age, gender, and comorbidities from study populations represented in cardiovascular device trials, 32 while the impact of therapeutics on patients with multiple chronic conditions in particular remains an important public health concern. 33 As the rigor of the local coverage decisions review has been called into question, 34 it may be worthwhile for CMS to lower its threshold for considering NCDs for high-risk devices, and to specifically focus on the need for evidence generation among populations similar to Medicare beneficiaries prior to providing coverage. CMS could then require device manufacturers to collect clinical data specifically pertaining to Medicare beneficiaries prior to issuing approval for reimbursement. Such a strategy would require funding to allow CMS to afford greater resources to conduct a larger number of NCDs. This strategy would alter the assumption that new high-risk devices could earn reimbursement through local coverage decisions alone. If more sponsors anticipate that their products will be subject to NCDs, this would likely incentivize their simultaneous engagement with FDA and CMS through parallel review. Conversely, if sponsors continue to assume that the risks of attracting scrutiny through an NCD remain small, they will have little to gain from encouraging additional oversight or proactively testing their devices in older, sicker patients.
The infrequent use of parallel review to date suggests that device sponsors' criticism of the successive FDA and CMS review processes and pleas for efficiency reflect a broader desire for minimizing regulatory risk. However, this program holds the potential to buttress regulatory standards of both agencies, requiring increased data collection for older patients not often included in clinical trials, while making potentially life-saving products available to the public more quickly. While time spent under review represents one element of risk to device manufacturers, sponsors also weigh potential restrictions on labeled indications and populations and post-market requirements when evaluating paths to market. These factors must be considered when motivating participation in parallel review, and may help the program achieve its goal of bringing useful medical devices to market, while also generating high-quality evidence specific to Medicare beneficiaries.