Abstract
Corneal decellularization has become an increasingly popular technique for generating scaffolds for corneal regeneration. Most decellularization procedures result in tissue swelling, thus limiting their application. Here, the use of a polysaccharide, dextran, to reduce swelling and conserve the native corneal structure during decellularization was investigated. Corneas were treated with 1% Triton X-100, 0.5% sodium dodecyl sulfate, and nucleases under constant rotation followed by extensive washing. To reduce swelling, decellularization solutions were supplemented with 5% dextran either throughout the whole decellularization process or during the washing cycles only. Quantitative analysis of DNA content showed a 96% reduction after decellularization regardless of the addition of dextran. Dextran resulted in a significant reduction in swelling from 3.85 ± 0.43 nm without to 1.94 ± 0.29–2.01 ± 0.37 nm (p < 0.05) remaining at similar dimensions to the native tissue (1.73 ± 0.23 nm). Tissue transparency was restored to all decellularized corneas following submersion in glycerol. Transmission electron microscopy (TEM) analysis found that dextran must be present throughout the decellularization protocol to preserve the native corneal architecture, anisotropy analysis demonstrated comparable results (0.22 ± 0.03) to the native cornea (0.24 ± 0.02), p > 0.05. Dextran can counteract the detrimental effects of decellularizing agents on the biomechanical properties of the tissue resulting in similar compressive moduli (mean before decellularization: 5.40 ± 1.18 kPa; mean after decellularization with dextran: 5.64 ± 1.34 kPa, p > 0.05). Cells remained viable in the presence of decellularized scaffolds. The findings of this study indicate that dextran not only prevents significant corneal swelling during decellularization but also enhances the maintenance of the native corneal ultrastructure.
Introduction
C
Tissue-engineering approaches rely on the fabrication of biomimetic corneal scaffolds using synthetic and naturally derived polymeric biomaterials.10–17 These scaffolds often fail to replicate the complex native structural and biochemical architecture of cornea. 18 Decellularized corneas overcome many of the limitations of other tissue-engineered methods since the inherent structural features are already present in the scaffold, they possess the required nanoscale topography and intrinsic biological cues reside within the matrix that support healthy cell phenotypes. 19 Decellularization involves removing nuclear and cellular material while preserving the structural components of the extracellular matrix (ECM). 20 Scaffolds derived by this method from a variety of tissues and organs have had clinical success,21,22 however, worldwide shortage of viable cadaveric corneal tissue has led to the exploration of alternative tissue sources.
Xenogenic porcine corneal scaffolds have attracted considerable interest in recent years,23–25 since they are inexpensive, readily available, and possess similar dimensional and refractive properties compared to human corneas. 8 Decellularization of porcine tissue has been achieved through multiple methods; however, as yet, a universal set of standardized protocols has not been outlined. A significant problem arising from decellularization processes is corneal swelling, limiting its potential use as a scaffold. Tissue swelling results in a scaffold with dimensions incompatible with the human eye and thus is unsuitable for transplantation. In addition, the highly organized ECM architecture that is essential to corneal transparency26,27 becomes disrupted after swelling and decellularization treatments leading to a reduction in the optical clarity of the cornea. Furthermore, ECM damage can alter the mechanical properties of the cornea. 28
The mechanical properties of the native corneas must be approximated in any successful corneal replacement, and is required to withstand internal and external forces that are applied to the eye, for example, intraocular and eyelid pressure and eye rubbing.29–31 Dextran is a complex polysaccharide routinely used at a concentration of 5% to act as a deswelling agent during organ culture to preserve human corneas32–35 although to date its use in controlling swelling during decellularization has not been investigated. Dextran increases colloidal osmotic pressure in cornea storage medium, limiting movement of water into the stroma by maintaining a high osmotic pressure. 34
In this study, we systematically evaluated the effect of dextran on swelling and ultrastructural maintenance of the cornea during decellularization.
Methods
All chemicals and reagents were purchased from Sigma-Aldrich, unless otherwise stated.
Dissection of porcine corneas
Adult porcine eyes were obtained from a local abattoir within 2 h postmortem. Eight millimeter corneal buttons were dissected and washed for 18 h in phosphate-buffered saline (PBS) containing 2% (v/v) antibiotic/antimycotic (0.02 μg/mL gentamycin, 0.5 ng/mL amphotericin B combination, Gibco; hereafter referred to as antibiotic/antimycotic wash solution).
Generation of an acellular porcine scaffold
Four corneal groups were tested: (i) agitated control, 2% antibiotic/antimycotic wash solution; (ii) detergent treatment, 0.5% (w/v) sodium dodecyl sulfate (SDS) and 1% (v/v) Triton X-100 in diH2O; (iii) detergent/dextran treatment, as detergent treatment with the addition of 5% (w/v) dextran throughout the procedure; (iv) detergent/dextran wash treatment, as detergent treatment followed by washing in 5% dextran. An additional control group, native cornea, represented freshly dissected corneas that did not undergo processing, except for brief washing in 2% antibiotic/antimycotic wash solution before analysis. All conditions were mechanically agitated on a gyratory rocker at 70 rpm for 72 h at room temperature (RT).
Corneas, except agitated control, were then treated with nucleases as previously described. 12 All corneas were then washed for 48 h with PBS containing 1% antibiotic/antimycotic. This was supplemented with 5% dextran for the detergent/dextran and detergent/dextran wash groups.
Scaffold swelling
Following decellularization, corneal swelling was analyzed. Images were recorded using a stereo microscope (Leica S6 D; Leica Microsystems) coupled to a digital camera (Canon PowerShot S50 5MP; Canon). Corneal thickness was measured at the center of the tissue with a micrometer calliper (Silverline) using 12 individual samples for each condition.
Corneal transparency and light transmittance
To evaluate changes in transparency porcine corneas were imaged following decellularization. Subsequently, the corneas were submerged in 100% glycerol for 18 h to examine restoration of corneal clarity. Light transmittance was quantified before and after glycerol treatment. The absorbance of light at several different wavelengths ranging from 350 to 700 nm was determined with a microplate reader (BioTek™ Synergy HTX), as previously described. 12 The transmittance of light was calculated by inverting the absorbance reading. Ultrapure water was used as a baseline control. Images were also recorded of the letter A through the corneas to give a visual representation of the difference in transparency.
Biochemical analysis
Corneal buttons were freeze-dried (Christ-Alpha 1–4 LSC Freeze Dryer) and the dry weight was recorded. Corneas were digested in a papain solution (125 μg/mL) in 0.2 M sodium phosphate buffer, 5 mM disodium ethylenediaminetetraacetic acid, and 10 mM cysteine hydrochloride, pH 6.4 for 18 h at 65°C, as previously described. 36 DNA was extracted and purified using a DNeasy Blood and Tissue Kit (Qiagen) and was quantified with a Quant-iT™ PicoGreen® dsDNA Assay Kit (Molecular Probes) according to the manufacturer's instructions. Fluorescence was measured at an excitation wavelength of 480 nm and an emission wavelength of 520 nm (Tecan Infinite® 200 PRO).
Sulfated glycosaminoglycan (sGAG) content was determined using a 1,9 dimethyl methylene blue (DMMB) assay (Biocolor) according to the manufacturer's protocol. DMMB dye was added to papain-digested samples and mixed thoroughly. The solution was then centrifuged and the supernatant removed. The remaining pellet was resuspended in dissociation dye and the sGAG content was measured by reading the absorbance at 656 nm.
Collagen content was determined using a Sircol™ soluble collagen assay (Bicolor) according to the manufacturer's protocol. First, freeze-dried corneas were digested at 4°C in pepsin extraction reagent (0.1 mg/mL in 0.5 M acetic acid) for 18 h. Sircol dye reagent was added to pepsin-digested samples and mechanically agitated for 30 min before being centrifuged. The pellet was washed in ice-cold acid salt and centrifuged. The pellet was resuspended in Alkali reagent and collagen content was measured by reading the absorbance of the solution at 555 nm. Values for each assay were calculated by interpolation from a standard curve and normalized to dry weight.
Histological staining
Scaffolds were fixed in 4% paraformaldehyde for 18 h. Samples were then dehydrated through a graded series of alcohol from 70% to 100% followed by immersion in xylene. Samples were then embedded in paraffin wax and 5 μm sections were cut using a microtome (Leica RM2125; Leica Biosystems). Hematoxylin and eosin (H&E) staining was used to visualize cell presence, while 1% alcian blue 8GX solution and picrosirius red staining was performed to examine sGAG and collagen retention, respectively. All samples were imaged using an inverted light microscope coupled to a digital camera (Olympus BX 41; Olympus). For nuclear staining, slides were incubated with 4′, 6-diamidino-2-phenylindole (1:500; DAPI) and imaged using an inverted epifluorescent microscope (Olympus IX 71; Olympus).
Mechanical testing
Unconfined compression experiments were performed. Samples were mechanically tested between two flat impermeable platens connected to 5N load cell (Zwick Z005; Zwick Roell) as previously described. 37 Samples were preloaded to 10% strain and allowed to fully relax. The samples were then dynamically loaded at frequencies of 0.1 Hz and 1 Hz. The compressive modulus was calculated from the first loading cycle while the dynamic modulus was calculated from the average stress and strain amplitude values over five cycles.
Transmission electron microscopy
Decellularized scaffolds were fixed in 3% glutaraldehyde in sodium cacodylate buffer solution (0.2 M, pH 7.2) for 18 h at 4°C. The scaffolds were postfixed with 1% osmium tetroxide in sodium cacodylate buffer solution for 2 h at room temperature, followed by washing in sodium cacodylate buffer (0.1 M, pH 7.2), serial dehydrations in ethanol, and washing in propylene oxide (TAAB Laboratories Equipment Ltd.). The scaffolds were embedded in araldite resin (TAAB Laboratories Equipment Ltd.) and sectioned (Leica EM UC6; Leica Biosystems). Ultrathin sections (90 μm) were contrasted with uranyl acetate and lead citrate and observed on a transmission electron microscope (FEI Tecnai Biotwin T12), operating at 100 kV. Images were recorded using a SIS Megaview digital camera (Olympus).
Collagen fibril alignment was quantified using FibrilTool, an Image J (NIH) plug-in, as previously described. 38 This plug-in applies circular statistics to analyze variance in tangent direction establishing whether fibrils are well ordered. A score is applied, 0 for no order or isotropic fibrils and 1 for perfectly ordered or anisotropic fibrils. Five distinct regions were chosen in 3 or 4 images for each condition and the anisotropy of collagen fibrils was measured. Fibril diameter was determined using the line drawing tool in Image J and was measured relative to the length of a 1 μm scale bar.
Evaluation of scaffold biocompatibility
Human keratocyte-derived fibroblast cells were cultured in the presence of decellularized corneas to evaluate biocompatibility. Cells were isolated as previously described 39 in accordance with the Declaration of Helsinki and cultured in low glucose Dulbecco's modified Eagle's medium supplemented with 10% (v/v) fetal bovine serum (FBS; Hyclone), 100 U/mL penicillin (Gibco), 100 μg/mL streptomycin (Gibco), and 250 ng/mL amphotericin-B, hereafter called lgDMEM. 1 × 105 passage 4 cells were seeded into six-well plates (Grenier Bio-One) and allowed to adhere at 37°C for 18 h.
Corneal buttons were placed in six-well ThinCert™ culture inserts (Grenier Bio-One) with pore size of 1 μm above the cells. All scaffolds were cultured in lgDMEM. Fibroblasts alone formed the control group. In addition, cells were cultured in lgDMEM supplemented with 5% dextran, termed cells (DM), to examine its effect on cell viability. Cell viability was assessed at 0, 3, 7, and 14 days using a PrestoBlue™ assay (Molecular Probes) according to the manufacturer's instructions. Results were calculated relative to viability at day 0.
Statistics
Statistical analysis was performed using GraphPad Prism software (GraphPad Software, Inc.). All data are represented as mean ± standard deviation. The statistical differences were evaluated by either One-Way or Two-Way ANOVA, followed by Bonferroni's multiple comparisons tests. Significance was accepted at a level of p < 0.05.
Results
Scaffold swelling and hydration
All scaffolds underwent a degree of swelling compared to freshly dissected native corneas (Fig. 1A). The average thickness of native corneas was 1.73 ± 0.23 mm while the thickness of agitated control (9.59 ± 1.07 mm) and detergent only corneas (3.85 ± 0.43 mm) were significantly greater (Fig. 1B). Scaffolds decellularized using protocols containing dextran were more comparable to the native corneas with a mean thickness of 2.01 ± 0.37 mm for detergent/dextran and 1.94 ± 0.29 mm for detergent/dextran wash. Dextran-treated corneas were significantly thinner than scaffolds decellularized using detergent only.

Percentage swelling was also analyzed (Fig. 1C), the mean increase in swelling compared to the native cornea (100.00% ± 0.00%) for each treatment was as follows: 554.80% ± 61.98% for agitated control, 222.80% ± 24.89% for detergent both significantly greater than the native cornea, while detergent/dextran and detergent/dextran wash had increases of 116.39% ± 21.61% and 112.34% ± 16.20%, respectively, with no significant difference to the native tissue.
Light transmittance and corneal transparency
There was a significant difference in the transparency and light transmittance of native corneas and decellularized corneas immediately after treatment at all the wavelengths analyzed (Fig. 2A, B). In all cases, the treated corneas transmitted significantly less light than the native controls. Images recorded of the corneas demonstrate the difference in transparency before and after decellularization treatment. To examine the restoration of corneal clarity the scaffolds were soaked in 100% glycerol and light transmittance was examined after 18 h (Fig. 2C). All corneas, except the agitated control, transmitted significantly more light than the native cornea after 18 h. Detergent/dextran and detergent/dextran wash had similar levels of light transmittance to the native cornea that was treated with glycerol (+ Gly), with detergent/dextran showing no significant difference. The images show that all decellularized corneas regained a level of transparency (Fig. 2A).

Residual porcine DNA
H&E and DAPI analysis showed cells and nuclei present throughout the cornea before decellularization, while none remained after decellularization treatments (Fig. 3A). Cells were still visible in the agitated control group. Quantification of DNA revealed that irrespective of the decellularization protocol, there was a significant reduction of DNA compared to native corneas (Fig. 3B). The mean DNA content of native and agitated control corneas was 3,335.84 ± 343.27 and 1494.78 ± 100.76 ng/mg respectively, while the yield from decellularized scaffolds was 120.00 ± 53.28, 124.82 ± 20.40, and 111.61 ± 12.54 ng/mg for detergent, detergent/dextran, and detergent/dextran wash corneas. This technique resulted in an average reduction in DNA of 96%.

sGAG and collagen content following decellularization
Quantitative results revealed that all decellularization protocols caused a significant reduction in sGAG content compared to native corneas (Fig. 4A). Native and agitated controls contained 7.98 ± 0.22 and 6.59 ± 0.74 μg/mg sGAG respectively. Detergent alone retained 2.14 ± 0.15 μg/mg of sGAG while detergent/dextran and detergent/dextran wash had levels of 3.87 ± 0.45 and 2.92 ± 0.22 μg/mg respectively. Histological staining with alcian blue demonstrated that all decellularized corneas displayed significantly weaker staining for sGAG (Fig. 4C). The collagen content within the samples was maintained after decellularization with no significant difference from the control samples, all contained ∼0.8 mg/mg collagen (Fig. 4B). Histological staining with picrosirius red confirmed these findings (Fig. 4C).

Mechanical properties of decellularized corneas
To examine the effect of decellularization on the corneal mechanical properties dynamic and compressive loading tests were performed (Fig. 5). The compressive modulus of agitated control (2.04 ± 0.94 kPa), detergent (1.86 ± 0.85 kPa), and detergent/dextran wash (3.65 ± 1.16 kPa) corneas was significantly lower than the native cornea (5.40 ± 1.18 kPa), while detergent/dextran (5.64 ± 1.34 kPa) had a similar modulus (Fig. 5A).

The decellularization process caused a significant reduction in the dynamic modulus of the corneal stroma at 0.1 Hz of agitated control (3.83 ± 1.99 kPa) and detergent (6.71 ± 2.49 kPa) corneas compared to the native tissue (15.20 ± 4.38 kPa), while detergent/dextran wash was not significantly different at 14.90 ± 3.66 kPa (Fig. 5B). In contrast, detergent/dextran displayed a significantly higher modulus of 21.70 ± 2.45 kPa. There was a similar trend for 1 Hz dynamic modulus with agitated control (4.09 ± 2.55 kPa) and detergent (4.83 ± 1.37 kPa) being significantly lower than the native control (17.50 ± 5.13 kPa), while detergent/dextran wash displayed similar properties with a modulus of 19.00 ± 4.97 kPa (Fig. 5C). The dynamic modulus at 1 Hz for detergent/dextran was again significantly greater than the native cornea at 25.94 ± 2.88 kPa.
Ultrastructure of the cornea and collagen fibril alignment
Collagen fibril orientation and disruption to the corneal architecture following decellularization was examined using transmission electron microscopy (TEM) (Fig. 6). The typical orthogonal fibril arrangement was evident in the native cornea, with collagen fibrils tightly packed within the tissue. The agitated control had similar structure, although the collagen bundles were more widely spaced due to corneal swelling. Decellularization with detergents alone caused a loss of collagen fibril alignment. The use of dextran revealed a more uniform arrangement of fibrils, with detergent/dextran most resembling the native cornea.

Transmission electron microscopy images of control and decellularized corneas; images depict collagen lamellae banding patterns, scale bar = 2 μm, collagen fibril bundles, scale bar = 1 μm, detail of fibrils, scale bar = 0.5 μm.
Quantification of fibril alignment demonstrated a significant reduction in fibril alignment for detergent (0.14 ± 0.01) and detergent/dextran wash (0.15 ± 0.06) groups compared to the native control (0.24 ± 0.02) (Fig. 7A). Agitated control and detergent/dextran maintained similar fibril alignment to fresh stromal tissue with an anisotropy of 0.18 ± 0.04 and 0.22 ± 0.03 respectively. When fibril diameter was quantified (Fig. 7B), agitated control (28.78 ± 1.68 nm) and detergent/dextran (28.33 ± 1.23 nm) corneas had similar diameters compared to the native control (30.98 ± 2.46 nm). However, detergent (25.58 ± 1.15 nm) and detergent/dextran wash (26.41 ± 2.36 nm) displayed a small but significant reduction in diameter.

Scaffold biocompatibility
Cell viability assays demonstrated a small but significant increase in cell proliferation for all groups, except cells only and cells (DM), over the culture period of 14 days compared to day 0 (Fig. 8). There was an initial surge in cell proliferation for the first 3 days in the presence of corneal tissue. There was no significant difference in cell proliferation when comparing detergent and detergent/dextran to the agitated control at day 3, while detergent/dextran wash had significantly enhanced proliferation. By day 7 there was no significant difference between agitated control and detergent/dextran, however, both detergent and detergent/dextran wash demonstrated significantly greater proliferation. At day 14 cell proliferation for both dextran-treated corneas was similar to the agitated control, while that for detergent was significantly greater.

Percentage of viable cells relative to day 0 as measured by cell proliferation at day 0, 3, 7, and 14. (n = 6, ± standard deviation). Color images available online at
Despite the significance observed the values for cell viability for each cornea treatment were in a similar range to each other, between ∼105–115% cell viability over 14 days relative to day 0. Cells cultured in the presence of dextran showed a significant increase in cell proliferation at day 3 compared with cells alone, this continued to day 7 after which the cells behaved as in normal lgDMEM with no significant difference between the two by day 14.
Discussion
Decellularized scaffolds require that cellular and nuclear components of the tissue be removed while the composition and structure of the ECM are maintained.18,20 This approach to biologic scaffold generation has been successful with a variety of tissue types and the resultant acellular ECM has been examined in both preclinical studies and human clinical applications.22,40–42
One problem associated with corneal decellularization is control of tissue hydration.43–45 During an effective decellularization process the endothelium of the cornea is completely removed. The corneal endothelium is essential for maintaining human vision by regulating the hydration state of the stroma by way of Na+/K+-ATPase pump functioning.46–48 Once the cornea is denuded of the endothelium it undergoes extensive swelling.43–45 In this study, the use of dextran during decellularization was investigated to prevent swelling and maintain the discrete and complex features of the corneal ECM.
Dextran is routinely used to maintain the hydration state of corneas in storage via colloid osmotic pressure. 34 Before keratoplasty, a concentration of 5% dextran is routinely used to maintain graft thickness.35,49,50 The addition of dextran during corneal decellularization has only been previously used in conjunction with ultrahigh hydrostatic pressure (HHP).43,44 The authors in these studies used dextran at a concentration of 3.5%, although no rationale was provided for the use of this concentration. Sasaki et al. (2009) used dextran before decellularization to maintain the native corneas dimensions, however, the effect of dextran was not described. Hashimoto et al. (2010) used dextran to prevent swelling during the washing step to avoid the swelling observed during washing without dextran. Dextran was not used during the decellularization procedure itself. The examination of the effect of dextran was limited to swelling analysis that demonstrated a significant reduction in corneal thickness. There was no direct comparison of the addition of dextran to those without dextran in the decellularization process for other parameters analyzed. The primary emphasis of these studies was on decellularization with HHP and not the addition of dextran. However, this decellularization method is limited by the need for expensive specialized equipment capable of applying pressures up to 1 GPa to the corneas. 18
Our study used relatively inexpensive detergents and enzymes that are highly effective at removing cells and cellular debris. The addition of dextran during decellularization was successful in regulating scaffold water content. Corneas decellularized or washed in the presence of dextran displayed a marked decrease in swelling compared to those with detergents and nucleases alone, with dimensions remaining similar to the native cornea, consistent with HHP studies.43,44 The extensive swelling observed for the agitated control can be attributed to the physiological pH, pH 7.40, of the PBS solution used. The isoelectric point of the cornea is considered to be close to pH 4 and the stroma undergoes hydration and swelling when moving away from this value.51,52 The decellularization solutions used in this study were determined to be pH 4.00 and pH 4.07 for detergent and detergent/dextran, respectively. This suggests that dextran serves as an adjunct to the standard decellularization solution to prevent significant swelling.
The concentrations of detergents in our study, 0.5% SDS and 1% Triton-X 100, have previously been used to effectively decellularize the cornea.28,53–55 The use of detergents and nucleases for decellularization was effective in removing cellular components from the cornea and these results are consistent with other reports.23,53,54,56 The effectiveness of this method is generally considered to be due to adequate disruption of the ECM enabling effective infiltration of decellularization factors into the scaffold. 57 However, the addition of dextran restricts corneal swelling, minimizing ECM disruption. Despite this the decellularizing agents were able to infiltrate the scaffold effectively and remove cellular components.
Biologic ECM scaffolds are an attractive choice for tissue engineering and regenerative applications since the structure and composition of ECM is similar between species.57,58 Maintenance of these features will enhance the functionality of the scaffold although any protocol aimed at complete cellular removal will result in some ECM disruption.57,59,60 Furthermore, the ECM contains intrinsic biological signals, including growth factors and cytokines, which play an important role in the maintenance of cell migration, proliferation, and phenotype,19,31,61 and can affect the host remodeling response following in vivo implantation.58,59,61,62 The GAG content after decellularization provides a useful indication as to the extent that biochemical cues and ECM components have been lost from the cornea. The concentration of GAGs was reduced in corneas decellularized using detergents even with the addition of dextran. Optimization of the concentration of detergents may improve the retention of GAGs in the ECM.
In addition to biochemical cues, the ultrastructural and topographical characteristics of the tissue play an important role in modulating cell behavior. 62 In the case of the cornea, maintenance of collagen fibril alignment is essential for transparency of the replacement tissue. The predominant corneal collagen is collagen type I.27,63 The uniform diameter and high level of lateral order of these fibrils gives the cornea its transparency. 26
A simple method to evaluate the maintenance of this architecture, which is often employed, is to soak corneas in glycerol after decellularization20,56,64 and evaluate changes macroscopically and by measuring light transmittance. Transparency was enhanced in all corneas after soaking in glycerol; however, the dextran-treated corneas had a similar level of light transmittance to the native cornea that was also treated in glycerol suggesting that the histoarchitecture of the stroma was maintained for these groups, similar results were obtained in the HHP studies using dextran.43,44 In addition, collagen content was maintained in all groups regardless of the decellularization method used. This is in agreement with previous decellularization studies carried out on annulus fibrosis tissue, 65 tendon, 66 and articular cartilage. 67
Ultrastructural maintenance with TEM analysis demonstrated that scaffolds treated with detergent/dextran most closely resembled the native cornea. Our data showed that detergent alone had significantly enhanced light transmittance after glycerol treatment well above the level of the native tissue and greater than dextran-treated corneas, however, TEM analysis confirmed significant disruption to the collagen architecture. It has been demonstrated that glycerol can reduce light scattering and improve the optical transmittance of scaffolds, 68 however, in this case glycerol could have disguised the damage caused to the cornea by the detergent only decellularization process.
Fibril diameter plays an important role in corneal transparency. The corneal stroma consists of uniformly aligned collagen fibrils of 25–35 nm diameter.69,70 In this study, fibril diameter for treated samples ranged from 25.58 to 28.78 nm consistent with normal stromal diameters, despite the decrease observed for detergent and detergent/dextran wash groups. Collagen fibrils in the normal cornea are weak at scattering light as their diameter is smaller than the wavelength of light resulting in a transparent tissue. 71 Collagen fibril diameter is increased in corneal wound and scar tissue resulting in increased light scattering.72,73 Huang and Meek (1999) suggested that fibril diameter doesn't change much over a range of hydrations, this is consistent with the data presented in this study that there was little change in fibril diameter and thus fibril diameter has little impact on light transmittance in this study.
The corneal mechanical properties can be ascribed to the gross arrangement of collagen fibrils within multiple lamellae that lie approximately parallel to the tissue surface. 26 An engineered corneal replacement has to withstand suturing during surgery 74 in addition to membrane stresses caused by intraocular pressure and external forces such as eyelid pressure. 30 Many strategies to engineer replacement tissue involve the use of soft hydrogels that have mechanical properties inferior to the native tissue. 75 The elastic modulus of the bulk cornea has previously been estimated by several different methods resulting in a wide range of values, 5 kPa to 20 MPa.30,76–79 The modulus obtained depends on the testing method, testing conditions, and orientation of the cornea with respect to applied stress. 79 Hatami-Marbini and Etebu performed out-of-plane compressive testing on corneal tissue and found the compressive modulus could range between 0.6 and 13.8 kPa, and that the modulus increases with increasing strain,30,78 consistent with our findings.
Corneas treated with detergents alone had inferior mechanical properties compared to native corneas. This result correlates with a previous study that showed decellularization using SDS could reduce the ultimate tensile strength of the cornea from 4.5 to 2.5 MPa. 28 The addition of dextran enabled the decellularized scaffolds to retain the corneas mechanical characteristics. However, corneas treated with detergent/dextran resulted in an increased dynamic modulus ranging from 21–25 kPa compared to 15–17 kPa for the native corneas. Despite this increase the moduli determined are well within the limits of the normal cornea that have been determined previously in the literature. If a cornea is too weak it will be unusable as it will not be able to withstand the physiological loads experienced by the cornea. A cornea that is too stiff, however, can be detrimental to the functionality of a corneal replacement. Rigid artificial implants and keratoprostheses have experienced complications postimplantation including glaucoma, extrusion, and rejection. 80
The capacity of the scaffold to support host cells is essential when considering tissue-engineering applications.18,25,54 Viability assays revealed that cells remained viable in the presence of decellularized corneas and may even enhance cell proliferation compared to cells alone. It has been demonstrated that extract from the corneal stroma contains active factors that can enhance keratocyte proliferation, 81 which supports our findings. These results suggest that detergents added during the decellularization process that would have a detrimental effect on the cell viability are effectively removed by the washing process.
Conclusion
It has been demonstrated that the addition of dextran can be used to control swelling and reduce structural ECM damage during corneal decellularization. These findings should enable researchers who are developing decellularization protocols to significantly refine their technique and produce decellularized scaffolds that closely resemble the structure and composition of the native human cornea.
Footnotes
Acknowledgments
Research conducted with the financial support of Science Foundation Ireland and Marie-Curie Action COFUND grant number: 11/SIRG/B2104; EPSRC Engineering, Tissue Engineering and Regenerative Medicine (E-TERM), grant number: EP/I017801/1 and COST Action BM-1302, Joining Forces in Corneal Regeneration Research, as part of a short-term scientific mission. Technical assistance and expertise from Denise Mclean, in the Advanced Microscopy Unit (School of Life Sciences, University of Nottingham), is very much appreciated.
Disclosure Statement
No competing financial interests exist.
