Optimal coverage of prandial insulin requirements remains an elusive goal. The invention of rapid-acting insulin analogs (RAIAs) was a big step forward in reducing postprandial glycemic excursions in patients with diabetes in comparison with using regular human insulin; however, even with these, the physiological situation cannot be adequately mimicked. Developing ultrafast-acting insulins (UFIs)—showing an even more rapid onset of action and a shorter duration of action after subcutaneous (SC) administration—is another step forward in achieving this goal. The need for UFIs has been gradually recognized over the years, and subsequently, a number of different approaches to cover this need are in clinical development. A rapid increase in circulating insulin levels can be achieved by different measures: Modification of the primary structure of insulin molecule (as we know from RAIAs), addition of excipients that enhance the appearance in the monomeric state post-injection, or addition of enzymes that enable more free spreading of the insulin molecules in the SC tissue. Other measures to increase the insulin absorption rate increase the local blood flow nearby the insulin depot in the SC tissue, injecting the insulin intradermally or applying via another route, e.g., the lung. The development of these approaches is in different stages, from quite early stages to nearing market authorization. In time, daily practice will show if the introduction of UFIs will fulfill their clinical promise. In this review, the basic idea for UFIs will be presented and the different approaches will be briefly characterized.
Intradermal (ID) delivery has been shown to accelerate insulin pharmacokinetics (PK). We compared the PK and pharmacodynamic (PD) effects of insulin lispro administered before two daily standardized solid mixed meals (breakfast and lunch), using microneedle-based ID or traditional subcutaneous (SC) delivery.
The study included 22 subjects with type 1 diabetes in an eight-arm full crossover block design. One arm established each subject's optimal meal dose. In six additional arms, the optimal, higher, and lower doses (+30%, −30%) were each given ID and SC delivery, in random order. The final arm assessed earlier timing for the ID optimal dose (−12 versus −2 min). The PK/PD data were collected for 6 h following meals. Intravenous basal regular insulin was given throughout, and premeal blood glucose (BG) adjusted to 115 mg/dl.
The primary end point, postprandial time in range (70–180 mg/dl), showed no route-based differences with a high level of overall BG control for both SC and ID delivery. Secondary insulin PK end points showed more rapid ID availability versus SC across doses and meals (ΔTmax −16 min, ΔT50rising −7 min, ΔT50falling −30 min, all
This study indicates that ID insulin delivery is superior to SC delivery in speed of systemic availability and PK consistency and may improve postprandial glucose control.
In order to enhance the absorption of insulin following subcutaneous injection, excipients were selected to hasten the dissociation rate of insulin hexamers and reduce their tendency to reassociate postinjection. A novel formulation of recombinant human insulin containing citrate and disodium ethylenediaminetetraacetic acid (EDTA) has been tested in clinic and has a very rapid onset of action in patients with diabetes. In order to understand the basis for the rapid insulin absorption,
Analytical ultracentrifugation and dynamic light scattering were used to infer the relative distributions of insulin monomers, dimers, and hexamers in the formulations. Electrical resistance of the insulin solutions characterized the overall net surface charge on the insulin complexes in solution.
The results of these experiments demonstrate that the zinc chelating (disodium EDTA) and charge-masking (citrate) excipients used in the formulation changed the properties of RHI in solution, making it dissociate more rapidly into smaller, charge-masked monomer/dimer units, which are twice as rapidly absorbed following subcutaneous injection than RHI (Tmax 60 ± 43 versus 120 ± 70 min).
The combination of rapid dissociation of insulin hexamers upon dilution due to the zinc chelating effects of disodium EDTA followed by the inhibition of insulin monomer/dimer reassociation due to the charge-masking effects of citrate provides the basis for the ultra-rapid absorption of this novel insulin formulation.
Rapid-acting insulin analogs were introduced to the market in the 1990s, and these products have improved treatment of diabetes by shortening the optimum delay time between injections and meals. Compared with regular human insulin, rapid-acting insulin formulations also reduce postprandial glycemic excursions while decreasing risk of hypoglycemia. However, the current prandial products are not fast enough for optimum convenience or control.
Recombinant human hyaluronidase (rHuPH20) has been used to increase the dispersion and absorption of other injected drugs, and in the case of prandial insulin analogs, it confers both ultrafast absorption and action profiles. Animal toxicology studies have demonstrated excellent tolerability of rHuPH20, and human studies, involving over 60,000 injections of prandial insulin + rHuPH20 to date, have similarly shown excellent safety and tolerability. Studies using rapid-acting analog insulin with rHuPH20 have included clinic-based pharmacokinetic and glucodynamic euglycemic glucose clamp studies, test meal studies, and take-home treatment studies. Administration methods have included subcutaneous injection of coformulations of rapid-acting insulin + rHuPH20 as well as continuous subcutaneous infusion of coformulations or use of pretreatment of newly inserted infusion sets with rHuPH20 followed by standard continuous subcutaneous insulin infusion therapy.
These studies have demonstrated acceleration of insulin absorption and action along with improvement in postprandial glycemic excursions and reduction in hypoglycemia risks. Further, rHuPH20 reduces intrasubject variability of insulin absorption and action and provides greater consistency in absorption and action profiles over wear time of an infusion set. Further studies of rHuPH20 in the take-home treatment setting are underway.
Barriers to the use of prandial insulin regimens include inadequate synchronization of insulin action to postprandial plasma glucose excursions as well as a significant risk of hypoglycemia and weight gain. Technosphere® insulin (TI) is an inhaled ultra-rapid-acting human insulin that is quickly absorbed in the alveoli. With a time to maximum plasma drug concentration of approximately 14 min and a time to maximum effect of 35 to 40 min, TI more closely matches the postprandial insulin concentrations seen in nondiabetic individuals. Studies have shown that long-term administration of prandial TI in combination with long-acting basal insulin results in reductions in hemoglobin A1c comparable to conventional subcutaneously injected prandial insulins but with improved control of early postprandial BG. Furthermore, TI has been associated with less weight gain and a lower incidence of hypoglycemia, which may enhance patient satisfaction and acceptability of insulin therapy. This review discusses the clinical properties of TI and proposes strategies for optimal use.
The absorption profile of rapid-acting insulin analogs delivered subcutaneously is slow compared with physiological insulin. Shorter time to peak and shorter duration of insulin action are important steps toward reducing high postprandial blood glucose concentrations in diabetes therapy and are critical for the development of a closed-loop insulin delivery system. Many attempts have been made to develop more rapid-acting insulins. Since the 1950s, different approaches, such as jet injectors and sprinkler needles, which try to increase the absorption areas of injected insulin, have been developed; however, none of them are commonly used in diabetes therapy. Massage and heat increase tissue blood perfusion and, thereby, the absorption of subcutaneously applied insulin. The main focus of this article is a novel device that allows local application of heat to human skin. The device can be connected to a regular insulin pump. This device could demonstrate a significant effect on insulin absorption and postprandial glucose excursions in multiple clinical trials.
This review summarizes the clinical development of a family of ultra-rapid-acting recombinant human insulin formulations. These formulations use ethylenediaminetetraacetic acid (EDTA) to chelate zinc and thereby destabilize insulin hexamers. In addition, insulin monomer surface charges are chemically masked with citrate to prevent reaggregation. The first phase 1 trials were performed using BIOD-090, an acidic 25 unit U/ml insulin formulation, which contained disodium-EDTA (NaEDTA). When compared with regular human insulin (RHI) and/or insulin lispro in multiple phase 1 studies, BIOD-090 consistently showed more rapid absorption and/or onset of action. A standard meal challenge study also demonstrated improved postprandial glucose profiles associated with BIOD-090. However, increased patient exposure in larger phase 3 trials showed that this formulation was associated with an increased incidence of local injection site reactions, most commonly pain. A next generation formulation, BIOD-100, contained the same excipients as a standard insulin concentration of 100 U/ml. BIOD-100 maintained an ultra-rapid action profile and was associated with modest but significantly improved toleration when compared with BIOD-090. In order to further improve toleration, the hypothesis that NaEDTA contributed to discomfort by chelating endogenous calcium was tested by either substituting calcium-EDTA for NaEDTA or by adding calcium chloride to the NaEDTA formulation. These calcium formulations essentially eliminated the excess discomfort associated with BIOD-090 but were associated with less optimal pharmacokinetic profiles in humans. Recent efforts have succeeded in developing ultra-rapid-acting human insulin formulations with acceptable injection site toleration by optimizing concentrations of calcium (BIOD-125) and with the use of magnesium sulfate to mitigate discomfort (BIOD-123). Similar formulation technology has also been shown to accelerate absorption of insulin analogs in animal models.
Insulin therapy in youth with type 1 diabetes mellitus (T1DM) poses a special challenge because childhood is an unsteady state with increasing weight, height, and caloric needs, leading to varying insulin requirements. The current rapid-acting insulin analogs are not as fast and short-acting as needed to meet these challenges. This review describes the unique characteristics of insulin action in youth with T1DM based on previously published euglycemic clamp studies. It also explains the rationale behind the need for ultrafast-acting insulins to advance open- and closed-loop insulin therapy for the pediatric population with diabetes. Lastly, it briefly summarizes ongoing and future projects to accelerate insulin action in youth with T1DM.
The July 2012 issue of the
This article briefly describes 14 diabetes self-management video games, and, when available, cites research findings on their effectiveness. The games were found by searching the Health Games Research online searchable database, three bibliographic databases (ACM Digital Library, PubMed, and Social Sciences Databases of CSA Illumina), and the Google search engine, using the search terms “diabetes” and “game.” Games were selected if they addressed diabetes self-management skills.
Serious video games for health are designed to entertain while changing a specific health behavior. This article identifies behavioral principles that can guide the development of serious video games focused on changing a variety of health behaviors, including those attempting to decrease risk of obesity and type 2 diabetes. Guidelines discussed include how to develop video games that provide a solid foundation for behavior change by enhancing a player's knowledge and skill, ways in which personal mastery experiences can be incorporated into a video game environment, using game characters and avatars to promote observational learning, creating personalized experiences through tailoring, and the importance of achieving a balance between “fun-ness” and “seriousness.” The article concludes with suggestions for future research needed to inform this rapidly growing field.
Exergames (i.e., video games that require gross motor activity) may provide intrinsically motivating experiences that engage youth in sustained physical activity.
Thirty-one low-income 15– to 19-year-old overweight and obese African American adolescents were randomly assigned to a competitive exergame (
Compared with the competitive group, the cooperative players were significantly more intrinsically motivated to play (
Cooperative exergame play produced higher intrinsic motivation to play the exergame than competitive exergame play did. Intrinsic motivation that came from a desire for control/choice was related to higher energy expenditure during game play. Cooperative exergame play holds promise as a method for engaging overweight and obese youth in physical activity.
Physical inactivity is associated with obesity and type 2 diabetes. A key obstacle to physical activity is lack of motivation. Although some interactive exercise games (i.e., exergames—video games that require physical exertion in order to play) motivate players to exercise more, few games take advantage of group dynamics to motivate players' duration of exercise. In a test of the Kohler motivation gain effect, this study varied the ability level of a virtually presented partner in an interactive exergame that focused on abdominal strength to identify effects on a subject's (S') persistence with the task.
Male (
Using planned orthogonal contrasts on difference scores between blocks 1 and 2, results showed that persistence was significantly (
Virtually presented partners who are moderately more capable than participants are the most effective at improving persistence in exergame tasks.
This quasi-experimental study investigated a game intervention—specifically, an alternate reality game (ARG)—as a means to influence college students' physical activity (PA). An ARG is an interactive narrative that takes place in the real world and uses multiple media to reveal a story.
Three sections of a college health course (
A significant group
It is important that any intervention meet the needs and interests of its target population. Here, the ARG was designed in light of the learning preferences of today's college students—collaborative and social, experiential and media-rich. Our results provide preliminary evidence that a game intervention can positively influence PA within the college student population.
Some active video games have been found to promote physical activity adherence because of enjoyment. However, many active games are exercise themed, which may interfere with the distracting properties that make game-based exercise more enjoyable than traditional exercise. This study compared exercise-themed and game-themed active games to investigate differences in energy expenditure and enjoyment.
Young adults (
After adjustment for gender and weight, aerobic games produced 2.70 kcal/kg−1/h−1 (95% confidence interval 2.41, 3.00) greater energy expenditure than balance games (
Aerobic games produced greater energy expenditure but lower enjoyment than balance games, and a game-themed aerobic game was found more enjoyable than an exercise-themed aerobic game. Integrating more strenuous activity into entertaining games instead of games that simply simulate exercise may be a fruitful avenue for active game development.
This quasi-experimental exploratory study investigated neuropsychological effects of exercise among older adults with diabetes mellitus (DM) compared with adults without diabetes (non-DM), and it examined the feasibility of using a stationary bike exergame as a form of exercise for older adults with and without diabetes. It is a secondary analysis that uses a small dataset from a larger randomized clinical trial (RCT) called the Cybercycle Study, which compared cognitive and physiological effects of traditional stationary cycling versus cybercycling.
In the RCT and the secondary analysis, older adults living in eight independent living retirement facilities in the state of New York were enrolled in the study and assigned to exercise five times per week for 45 min per session (two times per week was considered acceptable for retention in the study) by using a stationary bicycle over the course of 3 months. They were randomly assigned to use either a standard stationary bicycle or a “cybercycle” with a video screen that displayed virtual terrains, virtual tours, and racing games with virtual competitors. For this secondary analysis, participants in the RCT who had type 2 DM (
Older adults with and without diabetes were able to use cybercycles successfully and complete the study, so the feasibility of this form of exercise for this population was supported. However, in contrast with the larger RCT, this small subset did not demonstrate statistically significant differences in executive function between the participants who used cybercycles and those who used stationary bikes with no games or virtual content on a video screen. Therefore, the study combined the two groups and called them “exercisers” and compared cognitive outcomes for DM versus non-DM patients. As predicted, exercisers with DM exhibited significant gains in executive function as measured by the Color Trails 2 test, controlling for age and education, while non-DM exercisers did not significantly gain in this measure [group x time interaction, F(1,16]) = 9.75;
These preliminary results support the growing literature that finds that exercise may improve cognition among older adult with DM. Additional research is needed to clarify why certain aspects of executive function might be differentially affected. The current findings may encourage physicians to prescribe exercise for diabetes management and may help motivate DM patients' compliance for engaging in physical activity.
The risk of acute hepatitis B among adults with diabetes mellitus is unknown. We investigated the association between diagnosed diabetes and acute hepatitis B.
Confirmed acute hepatitis B cases were reported in 2009–2010 to eight Emerging Infections Program (EIP) sites; diagnosed diabetes status was determined. Behavioral Risk Factor Surveillance System respondents residing in EIP sites comprised the comparison group. Odds ratios (ORs) comparing acute hepatitis B among adults with diagnosed diabetes versus without diagnosed diabetes were determined by multivariate logistic regression, adjusting for age, sex, and race/ethnicity, and stratified by the presence or absence of risk behaviors for hepatitis B virus (HBV) infection.
During 2009–2010, EIP sites reported 865 eligible acute hepatitis B cases among persons aged ≥23 years; 95 (11.0%) had diagnosed diabetes. Comparison group diabetes prevalence was 9.1%. Among adults without hepatitis B risk behaviors and with reported diabetes status, the OR for acute hepatitis B comparing adults with and without diabetes was 1.9 (95% confidence interval [CI] = 1.4, 2.6); ORs for adults ages 23–59 and ≥60 years were 2.1 (95% CI = 1.6, 2.8) and 1.5 (95% = CI 0.9, 2.5), respectively.
Diabetes was independently associated with an increased risk for acute hepatitis B among adults without HBV risk behaviors.
Most manufacturers of blood glucose monitoring equipment do not give advice regarding the use of their meters and strips onboard aircraft, and some airlines have blood glucose testing equipment in the aircraft cabin medical bag. Previous studies using older blood glucose meters (BGMs) have shown conflicting results on the performance of both glucose oxidase (GOX)- and glucose dehydrogenase (GDH)-based meters at high altitude. The aim of our study was to evaluate the performance of four new-generation BGMs at sea level and at a simulated altitude equivalent to that used in the cabin of commercial aircrafts.
Blood glucose measurements obtained by two GDH and two GOX BGMs at sea level and simulated altitude of 8000 feet in a hypobaric chamber were compared with measurements obtained using a YSI 2300 blood glucose analyzer as a reference method. Spiked venous blood samples of three different glucose levels were used. The accuracy of each meter was determined by calculating percentage error of each meter compared with the YSI reference and was also assessed against standard International Organization for Standardization (ISO) criteria. Clinical accuracy was evaluated using the consensus error grid method. The percentage (standard deviation) error for GDH meters at sea level and altitude was 13.36% (8.83%; for meter 1) and 12.97% (8.03%; for meter 2) with
There was variation in the number of time individual meters met the standard ISO criteria ranging from 72–100%. Results from all four meters at both sea level and simulated altitude fell within zones A and B of the consensus error grid, using YSI as the reference.
Overall, at simulated altitude, no differences were observed between the performance of GDH and GOX meters. Overestimation of blood glucose concentration was seen among individual meters evaluated, but none of the results obtained would have resulted in dangerous failure to detect and treat blood glucose errors or in giving treatment that was actually contradictory to that required.
To evaluate the feasibility of an implantable subconjunctival glucose monitoring system (SGMS) for glucose monitoring in humans, we investigated the
The new SGMS consists of an implantable ocular mini implant (OMI) and a hand-held fluorescence photometer. The implantable subconjunctival glucose sensor is composed of a fluorescence resonance energy transfer system based on Concanavalin A chemistry, embedded in a nelfilcon polymer hydrogel disk. Blood glucose changes in humans were induced by oral glucose intake and insulin injections.
The
For the first time, the possibility to measure glucose
The accuracy of continuous glucose monitoring (CGM) in non-critically ill hospitalized patients with heart failure or severe hyperglycemia (SH) is unknown.
Hospitalized patients with congestive heart failure (CHF) exacerbation (receiving IV or subcutaneous insulin) or SH requiring insulin infusion were compared to outpatients referred for retrospective CGM.
Forty-three patients with CHF, 15 patients with SH, and 88 outpatients yielded 470, 164, and 2150 meter–sensor pairs, respectively. Admission glucose differed (188 versus 509 mg/dl in CHF and SH,
The results suggest that SH and glycemic variability are more important determinants of CGM accuracy than known CHF status alone in hospitalized patients.
The concept of closed-loop control of glucose, in which continuous glucose sensing is coupled to a fully automated insulin delivery device, without human input, has been an attractive idea for diabetes management. This study presents a new class of hybrid nanogels that can integrate glucose sensing and glucose-responsive insulin release into a single nano-object.
Zinc oxide@poly[N-isopropylacrylamide (NIPAM)-acrylamide (AAm)- 2-aminomethyl-5-fluorophenylboronic acid (FPBA)] hybrid nanogels were synthesized and investigated for size, morphology, volume phase transition, photoluminescence properties, and
Zinc oxide@poly(NIPAM-AAm-FPBA) hybrid nanogels can sensitively and selectively detect glucose in highly reproducible fluorescent signals over the clinically relevant glucose concentration range of 18–540 mg/dl. The glucose-responsive volume phase transition of the nanogels can further regulate the release of the preloaded insulin. The insulin release from the nanogels exhibits the slowest rate (~5% released in 76 h) at a normal glucose level (108.0 mg/dl) but becomes quicker and quicker as the glucose increases to higher and higher levels.
The rationally designed hybrid nanogel can optically signal the glucose level with high sensitivity and selectivity and simultaneously regulate the insulin release rate in response to the glucose reading, which shows a promising concept toward the development of a miniaturized closed-loop glycemic control system.
Safe and effective closed-loop control (artificial pancreas) is the ultimate goal of insulin delivery. In this study, we examined the performance of a closed-loop control algorithm used for the overnight time period to safely achieve a narrow target range of blood glucose (BG) concentrations prior to breakfast. The primary goal was to compare the quality of algorithm control during repeated overnight experiments.
Twenty-three subjects with type 1 diabetes performed 2 overnight experiments on each of three visits at the study site, resulting in 138 overnight experiments. On the first evening, the subject's insulin therapy was applied; on the second, the insulin was delivered by an algorithm based on subcutaneous continuous glucose measurements (including meal control) until midnight. Overnight closed-loop control was applied between midnight and 6 a.m. based on hourly venous BG measurements during the first and second nights.
The number of BG values within the target range (90–150 mg/dl) increased from 52.9% (219 out of 414 measurements) during the first nights to 72.2% (299 out of 414 measurements) during the second nights (
Overnight controller performance improved when optimized initial control was given; this was suggested by the better metabolic control during the second night. Adequate controller run-in time seems to be important for achieving good overnight control. In addition, the findings demonstrate that hourly BG data are sufficient for the closed-loop control algorithm tested to achieve appropriate glycemic control.
Impaired dexterity has been reported to be prevalent in diabetes patients independent from the existence of diabetic neuropathy. This study was performed to investigate the impact of dexterity impairment on patient preference for two insulin pen injection devices (InnoLet and FlexTouch).
Ninety patients [54 male/36 female; age (mean ± standard deviation), 62 ± 8 years; disease duration, 18 ± 11 years; hemoglobin A1c, 7.2 ± 1.0%] were included in this investigation and were stratified into four different groups based on the results of a dexterity test (Jebsen—Taylor Hand Function Test) and assessment of visual impairment: 15 type 1 (group A) and 30 type 2 (group B) patients with impaired dexterity, 30 type 1/type 2 patients with visual impairment (group C), and 15 type 1/type 2 patients without any impairment (group D). The patients performed a cognitive function test (number connection test), were introduced to the devices in random order, and were asked to perform some mock injections before completing a six-item standardized preference questionnaire.
There was a strong preference for FlexTouch in all groups. All unimpaired patients (100%, group D) preferred FlexTouch, as did the vast majority in all other groups. Only 11% of the patients with impaired cognitive function preferred InnoLet, as did a few patients with more severely impaired dexterity or with visual impairment (group A, 13%; group B, 3%; group C, 14%).
Patient dexterity skills may have an influence on device preference, especially if the impairment is more pronounced.
Insulin therapy is the cornerstone of medical treatment for many patients with diabetes. Self-administration of this life-saving medication is no longer limited to the traditional vial and syringe. Instead, more and more patients worldwide are using prefilled insulin pen devices. Ease of use, convenience, confidence in dosing accuracy, and improved quality of life are just a few of their advantages. As with any medical technology, safe and proper use is vital. Many studies have examined pen device preference and usability. Until now, no study has included patients with both visual and dexterity impairments. To ensure safe and simple self-administration of insulin for all patients, it is time for the special needs of patients with diabetes to be considered not only during product development, but during postmarketing studies as well.
The clinical significance of blood glucose meter (BGM) error in the presence of increasing carbohydrate errors in diabetes patients who use both the BGM result and the carbohydrate estimation to dose insulin is unknown.
This Monte Carlo simulation modeled diabetes patients who calculate insulin dosages based on BGM results and carbohydrate estimations. It evaluated the likelihood of on-target insulin dosing and clinically significant insulin dose errors based on data from five BGMs with different levels of performance (expressed as bias and imprecision [coefficient of variation (%CV)]) and increasing levels of carbohydrate estimation errors. The study was performed across three separate preprandial glucose (PPG) ranges (90–150, 150–270, and 270–450 mg/dl).
When carbohydrate estimation is accurate (%CV = 0%), the likelihood for on-target insulin doses ranged 50.1–98.5%. The likelihood depended on BGM performance and PPG range. In the presence of carbohydrate estimation errors (%CV = 5–20%), the likelihood of on-target insulin dosages markedly decreased (range, 27.2–80.1%) for all BGMs, the likelihood of insulin underdosing (range, 0–12.8%) and overdosing (range, 0–32.3%) increased, and the influence of BGM error on insulin dosing accuracy was blunted. Even in the presence of carbohydrate error, the BGM with the best performance (bias 1.35% and %CV = 4.84) had the highest probability for on-target insulin dosages.
Both BGM and carbohydrate estimation error contribute to insulin dosing inaccuracies. The BGM with the best performance was associated with the most on-target insulin dosages.
Language barriers are significant hurdles for chronic disease patients in achieving self-management goals of therapy, particularly in settings where practitioners have limited nonprimary language skills, and in-person translators may not always be available. S-MINDS© (Speaking Multilingual Interactive Natural Dialog System), a concept-based speech translation approach developed by Fluential Inc., can be applied to bridge the technologic gaps that limit the complexity and length of utterances that can be recognized and translated by devices and has the potential to broaden access to translation services in the clinical settings.
The prototype translation system was evaluated prospectively for accuracy and patient satisfaction in underserved Spanish-speaking patients with diabetes and limited English proficiency and was compared with other commercial systems for robustness against degradation of translation due to ambient noise and speech patterns.
Accuracy related to translating the English-Spanish-English communication string from practitioner to device to patient to device to practitioner was high (97–100%). Patient satisfaction was high (means of 4.7–4.9 over four domains on a 5-point Likert scale). The device outperformed three other commercial speech translation systems in terms of accuracy during fast speech utterances, under quiet and noisy fluent speech conditions, and when challenged with various speech disfluencies (i.e., fillers, false starts, stutters, repairs, and long pauses).
A concept-based English-Spanish speech translation system has been successfully developed in prototype form that can accept long utterances (up to 20 words) with limited to no degradation in accuracy. The functionality of the system is superior to leading commercial speech translation systems.
Self-monitoring of blood glucose (SMBG) implicates a number of handling steps with the meter and the lancing device. Numerous user errors can occur during SMBG, and each step adds to the complexity of use. This report compares the required steps to perform SMBG of one fully integrated (the second generation of the Accu-Chek® Mobile), three partly integrated (Accu-Chek Compact Plus, Ascensia® Breeze®2, and Accu-Chek Aviva), and six conventional (Bayer Contour®, Bayer Contour USB, BGStar™, FreeStyle Lite®, OneTouch® Ultra® 2, and OneTouch Verio™Pro) systems. The results show that the fully integrated system reduces the number of steps to perform SMBG. The mean decrease is approximately 70% compared with the other systems. We assume that a reduction of handling steps also reduces the risk of potential user errors and improves the user-friendliness of the system.
Since the first successful use of insulin in 1921 to treat diabetes at Toronto General Hospital, the major advances in development of the medication itself have taken place in parallel with equally significant developments in the means of delivery. Administration of insulin remains parenteral. This article reviews the main variants in prescription-available delivery technology: Vial and syringe, pen injector, needle-free injection, and continuous subcutaneous insulin infusion pumps. For each of these, the background and major milestones are covered briefly and followed by a discussion of the latest product innovations, technologies, and implementations, which are all considered in the context of the interaction with users. The article concludes by reflecting upon how the progress in the technology of diabetes management can best serve the patient. The spectacular technological advances in medication, monitoring, and delivery since 1922 have transformed the lives of millions. However, the fact that we can add sophisticated technology to delivery devices and accessories does not mean it is always the best thing for the patient. Electronic sophistication may be welcomed by a young, eager type 1 diabetes patient, while a senior citizen who discovers he has type 2 diabetes may yearn for simplicity. Technology continues to provide great solutions, but the type of solution delivered must be matched to the user if the maximum benefit is to be achieved for all.
Continuous subcutaneous insulin infusion from an insulin pump depends on reliable transfer of the pumped insulin to the subcutaneous insulin depot by means of an insulin infusion set (IIS). Despite their widespread use, the published knowledge about IISs and related issues regarding the impact of placement and wear time on insulin absorption/insulin action is relatively small. We also have to acknowledge that our knowledge is limited with regard to how often patients encounter issues with IISs. Reading pump wearer blogs, for instance, suggests that these are a frequent source of trouble. There are no prospective clinical studies available on current IIS and insulin formulations that provide representative data on the type and frequency of issues with infusion sets. The introduction of new IISs and patch pumps may foster a reassessment of available products and of patient problems related to their use. The aim of this review is to summarize the current knowledge and recommendations about IISs and to highlight potential directions of IIS development in order to make insulin absorption safer and more efficient.
Cardiac autonomic neuropathy (CAN) is a serious complication of longstanding diabetes and is associated with an increased morbidity and reduced quality of life in patients with diabetes. The present study evaluated the prevalence of CAN diagnosed by reduced heart rate variability (HRV) using a newly developed device in a large, unselected, hospital-based population of patients with diabetes.
The study examined 323 patients consisting of 206 type 1 diabetes (T1DM) patients and 117 type 2 diabetes (T2DM) patients. The new handheld prototype Vagus™ was used to screen for CAN. Three different standardized cardiac reflex tests were performed to calculate HRV: 30:15 ratio, E:I ratio, and the Valsalva maneuver. An abnormal HRV in one test is indicative of early CAN, and if two or more tests show abnormal HRV, the diagnosis of CAN is established.
In total, 86% of examined patients completed all three tests. Each test was completed by more than 90% of the patients. The prevalence of established CAN was 23%, whereas 33% of the patients had early signs of CAN. The prevalence was higher in T2DM patients (27.8%) than in T1DM patients (20.6 %),
Cardiac autonomic neuropathy is frequent in both T2DM and T1DM patients, especially in those with other late diabetes complications. Screening for CAN with the new device is feasible.
Since 2000, there has been an ongoing debate regarding tightness of glycemic control in critically ill patients. An increased risk of hypoglycemia is observed in patients treated with an intensive insulin protocol targeting “normoglycemia,” probably accounting for a reduction of the overall benefit. Hypoglycemia is associated with neurological side effects and is found to be an independent predictor of mortality in most trials; however, long-term sequelae are rare if glucose is administered early. We describe a case of prolonged, extreme hypoglycemia in a critically ill patient treated according to an intensive insulin protocol who recovered without any neurological deficit at discharge.
In times of short health care budgets, reimbursement for self-monitoring of blood glucose (SMBG) in diabetes patients without insulin treatment is subject to debate. The Structured Testing Program (STeP) trial found a positive correlation of test frequency and improved hemoglobin A1c (HbA1c) levels in poorly controlled type 2 diabetes patients not treated with insulin.
A structured literature search for other clinical studies reporting on SMBG frequency was performed.
There is scarce evidence: Three trials, including STeP, noted a significant and relevant correlation between testing frequency and improved HbA1c levels (FA effect), whereas two studies did not. The comparability between the identified studies is problematic.
Future research should consider correlations between testing frequency and level of glycemic control. More emphasis should be placed on a structured approach to use SMBG and to address adherence to testing and therapy.